Prusa3D Prusament PP CF Black
MaterialID MAT0036

3DPIceland Labs
Generated with 3DPIceland Engineering Platform v70.0.4 DIRECT-IMPACT-METHODOLOGY - Izod and Charpy Methodology
2026-10-04 18:10:39
Measurement provenance
Canonical SQLite measured dates use ISO format; missing dates remain Not recorded.
| Module | Measured date |
|---|---|
| Tensile | Not recorded |
| Impact | Not recorded |
| Izod | 2026-10-02 |
| Charpy | 2026-10-02 |
| Stiffness | Not recorded |
Complete
6/6
41
Approved
Rigid specimen and result quality
| Module | Orientation | Average | Std. dev. | CV | Specimens | Confidence | Validation |
|---|---|---|---|---|---|---|---|
| Tensile | Upright | 7,644 MPa | 0,992 MPa | 0,13% | 10 | 10/10 | OK |
| Tensile | Flat | 27,224 MPa | 3,859 MPa | 0,142% | 10 | 10/10 | OK |
| Impact | Upright | 7,277 kJ/m² | 1,352 kJ/m² | 0,186% | 10 | 10/10 | OK |
| Impact | Flat | 20,461 kJ/m² | 1,441 kJ/m² | 0,07% | 10 | 10/10 | OK |
| Stiffness | Three-point bend | 1387,31 MPa | n/a | n/a | 1 | Complete | OK |
Recorded native inputs
| Module | Input set | Recorded values |
|---|---|---|
| Tensile | Upright force inputs (N) | 46, 54, 58, 42, 53, 39, 52, 46, 58, 49 |
| Tensile | Flat force inputs (N) | 156, 158, 172, 172, 167, 168, 175, 178, 179, 245 |
| Impact | Upright needle inputs (%) | 7,00, 9,00, 8,00, 7,00, 10,00, 10,00, 12,00, 10,50, 8,00, 8,00 |
| Impact | Flat needle inputs (%) | 27,00, 25,00, 26,00, 27,00, 23,00, 25,00, 26,50, 22,00, 23,00, 25,00 |
| Stiffness | Revolutions / degrees | 3 / 250 |
Approved test notes
Method and equipment context
Testing methodology · Engineering methodology whitepaper · Download PDF
Izod and Charpy Impact Strength
Izod Impact Strength · 2026-10-02
Instrument readings in kJ/m²
Mean 11.868 kJ/m²; Std. Dev 0.462 kJ/m²; CV 3.89%; Samples 10; Confidence 10
No break 0; invalid 0; excluded 0.
Comparative measurements; standard conformity unverified.
Charpy Impact Strength · 2026-10-02
Instrument readings in kJ/m²
Mean 12.017 kJ/m²; Std. Dev 0.194 kJ/m²; CV 1.617%; Samples 10; Confidence 10
No break 0; invalid 0; excluded 0.
Comparative measurements; standard conformity unverified.