Public test session report

Prusa3D Prusament Galaxy Black PLA Black

MaterialID MAT0112

3DPIceland Labs
Generated with 3DPIceland Engineering Platform v70.0.4 DIRECT-IMPACT-METHODOLOGY - Izod and Charpy Methodology
2026-10-04 18:10:39

Traceability boundary: canonical per-module measured dates are shown below. Session identifier, time of day, operator, printer/slicer profile and environmental conditions are Not recorded and are never inferred.

Measurement provenance

Canonical SQLite measured dates use ISO format; missing dates remain Not recorded.

ModuleMeasured date
TensileNot recorded
ImpactNot recorded
Izod2026-09-30
Charpy2026-09-30
StiffnessNot recorded
Rigid test status
Complete
Rigid result modules
6/6
Rigid specimen/result records
41
Public details
Approved

Rigid specimen and result quality

ModuleOrientationAverageStd. dev.CVSpecimensConfidenceValidation
TensileUpright19,964 MPa7,482 MPa0,375%1010/10OK
TensileFlat40,959 MPa9,294 MPa0,227%1010/10OK
ImpactUpright7,111 kJ/m²1,068 kJ/m²0,15%1010/10OK
ImpactFlat15,79 kJ/m²1,016 kJ/m²0,064%1010/10OK
StiffnessThree-point bend2883,003 MPan/an/a1CompleteOK

Recorded native inputs

ModuleInput setRecorded values
TensileUpright force inputs (N)153, 76, 188, 158, 49, 184, 103, 96, 115, 176
TensileFlat force inputs (N)334, 269, 315, 205, 192, 322, 317, 207, 192, 310
ImpactUpright needle inputs (%)10,50, 9,00, 7,50, 8,00, 7,50, 8,00, 11,00, 8,00, 10,00, 8,00
ImpactFlat needle inputs (%)19,00, 20,00, 19,00, 20,00, 17,50, 17,00, 20,00, 19,00, 20,00, 21,00
StiffnessRevolutions / degrees1 / 280

Approved test notes

Method and equipment context

Purpose-built tensile, pendulum impact and three-point bend apparatus are used with the governed Settings Manager constants consumed by ResultsService. Results are comparative and do not claim accredited-laboratory equipment.

Testing methodology · Engineering methodology whitepaper · Download PDF

Izod and Charpy Impact Strength

Izod Impact Strength · 2026-09-30
Instrument readings in kJ/m²
Mean 3.783 kJ/m²; Std. Dev 0.109 kJ/m²; CV 2.873%; Samples 10; Confidence 10
No break 0; invalid 0; excluded 0.
Comparative measurements; standard conformity unverified.

Charpy Impact Strength · 2026-09-30
Instrument readings in kJ/m²
Mean 4.659 kJ/m²; Std. Dev 1.86 kJ/m²; CV 39.918%; Samples 10; Confidence 10
No break 0; invalid 0; excluded 0.
Comparative measurements; standard conformity unverified.