Public test session report
Prusa3D Buddy3D Pearl Red PLA Red
MaterialID MAT0138

3DPIceland Labs
Generated with 3DPIceland Engineering Platform v70.0.4 DIRECT-IMPACT-METHODOLOGY - Izod and Charpy Methodology
2026-10-04 18:10:39
Traceability boundary: canonical per-module measured dates are shown below. Session identifier, time of day, operator, printer/slicer profile and environmental conditions are Not recorded and are never inferred.
Measurement provenance
Canonical SQLite measured dates use ISO format; missing dates remain Not recorded.
| Module | Measured date |
|---|---|
| Tensile | Not recorded |
| Impact | Not recorded |
| Izod | Not recorded |
| Charpy | Not recorded |
| Stiffness | Not recorded |
Rigid test status
Partial
Partial
Rigid result modules
4/6
4/6
Rigid specimen/result records
41
41
Public details
Approved
Approved
Rigid specimen and result quality
| Module | Orientation | Average | Std. dev. | CV | Specimens | Confidence | Validation |
|---|---|---|---|---|---|---|---|
| Tensile | Upright | 14,166 MPa | 4,868 MPa | 0,344% | 10 | 10/10 | OK |
| Tensile | Flat | 41,421 MPa | 5,549 MPa | 0,134% | 10 | 10/10 | OK |
| Impact | Upright | 7,277 kJ/m² | 1,459 kJ/m² | 0,2% | 10 | 10/10 | OK |
| Impact | Flat | 10,166 kJ/m² | 1,24 kJ/m² | 0,122% | 10 | 10/10 | OK |
| Stiffness | Three-point bend | 3324,545 MPa | n/a | n/a | 1 | Complete | OK |
Recorded native inputs
| Module | Input set | Recorded values |
|---|---|---|
| Tensile | Upright force inputs (N) | 72, 66, 118, 142, 102, 57, 103, 131, 52, 78 |
| Tensile | Flat force inputs (N) | 249, 242, 297, 247, 283, 239, 210, 306, 308, 312 |
| Impact | Upright needle inputs (%) | 7,00, 9,00, 11,00, 7,00, 7,00, 9,00, 12,00, 8,00, 10,50, 9,00 |
| Impact | Flat needle inputs (%) | 10,00, 12,00, 10,00, 13,00, 13,00, 14,00, 12,00, 14,00, 14,00, 12,50 |
| Stiffness | Revolutions / degrees | 1 / 195 |
Approved test notes
Method and equipment context
Purpose-built tensile, pendulum impact and three-point bend apparatus are used with the governed Settings Manager constants consumed by ResultsService. Results are comparative and do not claim accredited-laboratory equipment.
Testing methodology · Engineering methodology whitepaper · Download PDF