Public test session report

Thorwell Standard PLA Yellow

MaterialID MAT0142

3DPIceland Labs
Generated with 3DPIceland Engineering Platform v60.0.6 WORKFLOW-RELIABILITY - Bug Fixes and Workflow Reliability
2026-08-25 19:04:35

Traceability boundary: canonical per-module measured dates are shown below. Session identifier, time of day, operator, printer/slicer profile and environmental conditions are Not recorded and are never inferred.

Measurement provenance

Canonical SQLite measured dates use ISO format; missing dates remain Not recorded.

ModuleMeasured date
TensileNot recorded
ImpactNot recorded
StiffnessNot recorded
Status
Complete
Result modules
3/3
Specimen/result records
40
Public details
Approved

Specimen and result quality

ModuleOrientationAverageStd. dev.CVSpecimensConfidenceValidation
TensileUpright11,182 MPa4,892 MPa0,437%1010/10OK
TensileFlat44,743 MPa8,64 MPa0,193%1010/10OK
ImpactUpright113,546 kJ/m²19,315 kJ/m²0,17%1010/10OK
ImpactFlat215,77 kJ/m²41,446 kJ/m²0,192%99/10OK
StiffnessThree-point bend2493,408 MPan/an/a1CompleteOK

Recorded native inputs

ModuleInput setRecorded values
TensileUpright force inputs (N)82, 78, 108, 41, 110, 36, 34, 68, 118, 52
TensileFlat force inputs (N)337, 346, 313, 360, 183, 302, 274, 292, 214, 288
ImpactUpright needle inputs (%)5,00, 5,50, 7,00, 7,00, 5,00, 7,50, 7,50, 7,50, 7,50, 8,00
ImpactFlat needle inputs (%)11,00, 10,00, 10,00, 12,00, 13,00, 15,00, 16,00, 16,00, 11,50,
StiffnessRevolutions / degrees2 / 20

Approved test notes

Method and equipment context

Purpose-built tensile, pendulum impact and three-point bend apparatus are used with the governed Settings Manager constants consumed by ResultsService. Results are comparative and do not claim accredited-laboratory equipment.

Testing methodology · Engineering methodology whitepaper · Download PDF