Public test session report
Winkle Standard Winkle Purple PLA HD Purple
MaterialID MAT0162

3DPIceland Labs
Generated with 3DPIceland Engineering Platform v70.0.4 DIRECT-IMPACT-METHODOLOGY - Izod and Charpy Methodology
2026-10-04 18:10:39
Traceability boundary: canonical per-module measured dates are shown below. Session identifier, time of day, operator, printer/slicer profile and environmental conditions are Not recorded and are never inferred.
Measurement provenance
Canonical SQLite measured dates use ISO format; missing dates remain Not recorded.
| Module | Measured date |
|---|---|
| Tensile | Not recorded |
| Impact | Not recorded |
| Izod | Not recorded |
| Charpy | Not recorded |
| Stiffness | Not recorded |
Rigid test status
Partial
Partial
Rigid result modules
4/6
4/6
Rigid specimen/result records
41
41
Public details
Approved
Approved
Rigid specimen and result quality
| Module | Orientation | Average | Std. dev. | CV | Specimens | Confidence | Validation |
|---|---|---|---|---|---|---|---|
| Tensile | Upright | 13,504 MPa | 6,292 MPa | 0,466% | 10 | 10/10 | OK |
| Tensile | Flat | 41,236 MPa | 8,177 MPa | 0,198% | 10 | 10/10 | OK |
| Impact | Upright | 6,33 kJ/m² | 0,868 kJ/m² | 0,137% | 10 | 10/10 | OK |
| Impact | Flat | 11,366 kJ/m² | 1,151 kJ/m² | 0,101% | 10 | 10/10 | OK |
| Stiffness | Three-point bend | 3000,199 MPa | n/a | n/a | 1 | Complete | OK |
Recorded native inputs
| Module | Input set | Recorded values |
|---|---|---|
| Tensile | Upright force inputs (N) | 68, 88, 13, 59, 98, 101, 130, 52, 118, 151 |
| Tensile | Flat force inputs (N) | 226, 265, 320, 266, 216, 321, 239, 324, 178, 326 |
| Impact | Upright needle inputs (%) | 7,50, 8,50, 5,00, 8,00, 8,00, 8,00, 8,00, 8,00, 9,00, 8,00 |
| Impact | Flat needle inputs (%) | 13,00, 14,50, 11,50, 13,00, 12,00, 15,00, 15,00, 15,00, 15,00, 15,00 |
| Stiffness | Revolutions / degrees | 1 / 255 |
Approved test notes
Method and equipment context
Purpose-built tensile, pendulum impact and three-point bend apparatus are used with the governed Settings Manager constants consumed by ResultsService. Results are comparative and do not claim accredited-laboratory equipment.
Testing methodology · Engineering methodology whitepaper · Download PDF