Public test session report

Eryone Standard White PP White

MaterialID MAT0194

3DPIceland Labs
Generated with 3DPIceland Engineering Platform v60.0.6 WORKFLOW-RELIABILITY - Bug Fixes and Workflow Reliability
2026-08-25 19:04:35

Traceability boundary: canonical per-module measured dates are shown below. Session identifier, time of day, operator, printer/slicer profile and environmental conditions are Not recorded and are never inferred.

Measurement provenance

Canonical SQLite measured dates use ISO format; missing dates remain Not recorded.

ModuleMeasured date
TensileNot recorded
ImpactNot recorded
StiffnessNot recorded
Status
Partial
Result modules
2/3
Specimen/result records
40
Public details
Approved

Specimen and result quality

ModuleOrientationAverageStd. dev.CVSpecimensConfidenceValidation
TensileUpright8,321 MPa2,173 MPa0,261%1010/10OK
TensileFlat15,165 MPa0,976 MPa0,064%1010/10OK
ImpactUpright223,108 kJ/m²25,943 kJ/m²0,116%1010/10OK
ImpactFlat600,643 kJ/m²68,51 kJ/m²0,114%1010/10OK
StiffnessThree-point bendn/an/an/a0OK

Recorded native inputs

ModuleInput setRecorded values
TensileUpright force inputs (N)67, 66, 58, 69, 57, 54, 24, 53, 36, 57
TensileFlat force inputs (N)100, 98, 102, 100, 101, 101, 99, 81, 103, 101
ImpactUpright needle inputs (%)13, 12,5, 10, 12, 13,5, 15, 13, 13, 14,5, 15
ImpactFlat needle inputs (%)34, 32, 32, 37, 35, 37, 37, 47, 33, 33
StiffnessRevolutions / degrees

Approved test notes

Method and equipment context

Purpose-built tensile, pendulum impact and three-point bend apparatus are used with the governed Settings Manager constants consumed by ResultsService. Results are comparative and do not claim accredited-laboratory equipment.

Testing methodology · Engineering methodology whitepaper · Download PDF