Siraya Tech Fiberheart Black ABS CF Black
MaterialID MAT0199

3DPIceland Labs
Generated with 3DPIceland Engineering Platform v70.0.4 DIRECT-IMPACT-METHODOLOGY - Izod and Charpy Methodology
2026-10-04 18:10:39
Measurement provenance
Canonical SQLite measured dates use ISO format; missing dates remain Not recorded.
| Module | Measured date |
|---|---|
| Tensile | 2024-01-01 |
| Impact | Not recorded |
| Izod | 2026-10-01 |
| Charpy | 2026-10-01 |
| Stiffness | Not recorded |
Complete
6/6
40
Approved
Rigid specimen and result quality
| Module | Orientation | Average | Std. dev. | CV | Specimens | Confidence | Validation |
|---|---|---|---|---|---|---|---|
| Tensile | Upright | 3,264 MPa | 1,77 MPa | 0,542% | 9 | 9/10 | OK |
| Tensile | Flat | 31,561 MPa | 3,414 MPa | 0,108% | 10 | 10/10 | OK |
| Impact | Upright | 7,687 kJ/m² | 0,712 kJ/m² | 0,093% | 10 | 10/10 | OK |
| Impact | Flat | 15,254 kJ/m² | 0,888 kJ/m² | 0,058% | 10 | 10/10 | OK |
| Stiffness | Three-point bend | 2527,565 MPa | n/a | n/a | 1 | Complete | OK |
Recorded native inputs
| Module | Input set | Recorded values |
|---|---|---|
| Tensile | Upright force inputs (N) | 10, 36, 10, 13, 7, 37, 26, 25, 27, |
| Tensile | Flat force inputs (N) | 236, 203, 206, 213, 206, 218, 158, 179, 223, 210 |
| Impact | Upright needle inputs (%) | 9,5, 10, 10, 9,5, 8, 10, 10,5, 9, 8, 10 |
| Impact | Flat needle inputs (%) | 20, 18, 17, 18, 18, 20, 20, 18, 18, 19 |
| Stiffness | Revolutions / degrees | 2 / 10 |
Approved test notes
| Tensile | test |
Method and equipment context
Testing methodology · Engineering methodology whitepaper · Download PDF
Izod and Charpy Impact Strength
Charpy Impact Strength · 2026-10-01
Instrument readings in kJ/m²
Mean 7.828 kJ/m²; Std. Dev 0.711 kJ/m²; CV 9.088%; Samples 5; Confidence 5
No break 0; invalid 0; excluded 0.
Comparative measurements; standard conformity unverified.
Izod Impact Strength · 2026-10-01
Instrument readings in kJ/m²
Mean 7.167 kJ/m²; Std. Dev 0.3 kJ/m²; CV 4.18%; Samples 9; Confidence 9
No break 0; invalid 0; excluded 0.
Comparative measurements; standard conformity unverified.