Public test session report

Siraya Tech Fiberheart Black ABS Core CF Black

MaterialID MAT0207

3DPIceland Labs
Generated with 3DPIceland Engineering Platform v70.0.4 DIRECT-IMPACT-METHODOLOGY - Izod and Charpy Methodology
2026-10-04 18:10:39

Traceability boundary: canonical per-module measured dates are shown below. Session identifier, time of day, operator, printer/slicer profile and environmental conditions are Not recorded and are never inferred.

Measurement provenance

Canonical SQLite measured dates use ISO format; missing dates remain Not recorded.

ModuleMeasured date
Tensile2026-07-30
Impact2026-07-30
Izod2026-09-30
Charpy2026-09-30
Stiffness2026-07-30
Rigid test status
Complete
Rigid result modules
6/6
Rigid specimen/result records
41
Public details
Approved

Rigid specimen and result quality

ModuleOrientationAverageStd. dev.CVSpecimensConfidenceValidation
TensileUpright4,399 MPa1,966 MPa0,447%1010/10OK
TensileFlat34,069 MPa7,243 MPa0,213%1010/10OK
ImpactUpright8,801 kJ/m²0,399 kJ/m²0,045%1010/10OK
ImpactFlat17,109 kJ/m²0,868 kJ/m²0,051%1010/10OK
StiffnessThree-point bend3844,005 MPan/an/a1CompleteOK

Recorded native inputs

ModuleInput setRecorded values
TensileUpright force inputs (N)43, 25, 30, 23, 50, 34, 28, 13, 33, 7
TensileFlat force inputs (N)165, 296, 209, 183, 239, 201, 218, 253, 163, 288
ImpactUpright needle inputs (%)11, 10,5, 11, 11, 10, 11, 11, 11, 10, 11,5
ImpactFlat needle inputs (%)20, 22, 19,5, 20, 20, 22, 22, 22, 20, 21
StiffnessRevolutions / degrees1 / 120

Approved test notes

Method and equipment context

Purpose-built tensile, pendulum impact and three-point bend apparatus are used with the governed Settings Manager constants consumed by ResultsService. Results are comparative and do not claim accredited-laboratory equipment.

Testing methodology · Engineering methodology whitepaper · Download PDF

Izod and Charpy Impact Strength

Izod Impact Strength · 2026-09-30
Instrument readings in kJ/m²
Mean 7.414 kJ/m²; Std. Dev 0.151 kJ/m²; CV 2.037%; Samples 10; Confidence 10
No break 0; invalid 0; excluded 0.
Comparative measurements; standard conformity unverified.

Charpy Impact Strength · 2026-09-30
Instrument readings in kJ/m²
Mean 7.517 kJ/m²; Std. Dev 0.247 kJ/m²; CV 3.292%; Samples 10; Confidence 10
No break 0; invalid 0; excluded 0.
Comparative measurements; standard conformity unverified.